Primary stability in various levels of impacted implants: An ex vivo study


BİLMENOĞLU Ç., ÖZYURT A., ÇİLİNGİR A. A., TURAN F. N.

Journal of Oral Implantology, vol.47, no.5, pp.380-384, 2021 (SCI-Expanded, Scopus)

  • Publication Type: Article / Article
  • Volume: 47 Issue: 5
  • Publication Date: 2021
  • Doi Number: 10.1563/aaid-joi-d-19-00232
  • Journal Name: Journal of Oral Implantology
  • Journal Indexes: Science Citation Index Expanded (SCI-EXPANDED), Scopus, Biotechnology Research Abstracts, EMBASE, MEDLINE
  • Page Numbers: pp.380-384
  • Keywords: Dental implant, Partial impaction, Primary stability, RFA
  • Trakya University Affiliated: Yes

Abstract

The aim of this study was to examine the primary stabilization of different vertically impacted bone implants. Implant stability was measured by resonance frequency analysis. Forty-five dental implants were used and divided into 3 groups. Group 1 was placed 4 mm (1/3 impacted), group 2 was placed 8 mm (2/3 impacted), and group 3 was placed 12 mm (fully impacted). Implant stability quotient values were measured on the longitudinal and transversal axis by 2 independent researchers. The fully impacted group showed the significantly highest value among the groups (P, .05). There were statistically varying implant-stability quotient values between researchers. None of the 1/3-impacted implants’ value reached a 70 implant-stability quotient value.