Intermittent mode reversion to VOO of a dual chamber pacemaker model


ERDOĞAN O., ALTUN A., ÖZBAY G.

PACE - Pacing and Clinical Electrophysiology, cilt.24, sa.11, ss.1704-1705, 2001 (SCI-Expanded, Scopus) identifier identifier identifier

  • Yayın Türü: Makale / Tam Makale
  • Cilt numarası: 24 Sayı: 11
  • Basım Tarihi: 2001
  • Doi Numarası: 10.1046/j.1460-9592.2001.01704.x
  • Dergi Adı: PACE - Pacing and Clinical Electrophysiology
  • Derginin Tarandığı İndeksler: Science Citation Index Expanded (SCI-EXPANDED), Scopus
  • Sayfa Sayıları: ss.1704-1705
  • Anahtar Kelimeler: Dual chamber pacemaker, Mode reversion, VOO
  • Trakya Üniversitesi Adresli: Evet

Özet

This report describes an asymptomatic patient with a DDDR pacemaker system (Inos Cls, Biotronik) that showed intermittent mode reversion to VOO and sometimes to VVI without a detectable source of electromagnetic interference and oversensing. Patients with this Biotronik model should be carefully assessed along with a surface ECG before placing the programming head.