Reference jitter values for the sternocleidomastoid muscle with concentric needle electrodes


Kula E., Tanridag T., Kahraman Koytak P., Uluc K.

MUSCLE & NERVE, cilt.63, sa.1, ss.116-119, 2021 (SCI-Expanded, Scopus)

  • Yayın Türü: Makale / Tam Makale
  • Cilt numarası: 63 Sayı: 1
  • Basım Tarihi: 2021
  • Doi Numarası: 10.1002/mus.27094
  • Dergi Adı: MUSCLE & NERVE
  • Derginin Tarandığı İndeksler: Science Citation Index Expanded (SCI-EXPANDED), Scopus, Academic Search Premier, BIOSIS, CINAHL, EMBASE, MEDLINE
  • Sayfa Sayıları: ss.116-119
  • Trakya Üniversitesi Adresli: Hayır

Özet

Background The aim of this study was to establish reference jitter values for the voluntary activated sternocleidomastoid (SCM) muscle using a concentric needle electrode (CNE). Methods The study included 39 healthy participants (20 female and 19 male) aged 18-77 y. Jitter was expressed as the mean consecutive difference (MCD) of 80-100 consecutive discharges. Filters were set at 1 and 10 kHz. The mean MCDs for all participants were pooled, and the mean value +2.5 SD was accepted as the upper limit for the mean MCD. The upper limit for individual MCD was calculated using +2.5 SD of the upper 10th percentile MCD for individual participants. Results Mean age of the participants was 45 +/- 14.5 y. Mean MCD was 16.20 +/- 2.23 mu s (range: 12-21 mu s), and the upper limit of normal for mean MCD was 21.8 mu s. The mean value for 823 individual jitters was 23.3 +/- 4.61 mu s (range: 6.6-36.9 mu s), and the upper limit of normal for each individual jitter was 34.6 mu s. Conclusions The present findings indicate that upper normal limit for mean MCD is 22 mu s and for individual data it is 35 mu s.